GENERAL
Function: Microprocessor-based
Insertion or In-Line Mass Flow Meters for gases
Flow Element: Constant
Temperature Thermal Mass Flow Element consists of
two 316 SS clad platinum-wound RTDs
Flow Rate Output Signal7: 4-20mA
isolated linear flow signal on all products. Pulsed
outputs optional8
Temperature Output Signal7: 4-20mA
isolated linear temperature signal from 40°F
to 200°F (-40°F to 750°F optional) is
second output on SIG & SRG Series;menu selectable
in lieu of flow outputs on SIE & SRE Series.A
second temperature output is also available on the
SIL and SRL Series, but it is 0–5VDC linear
instead of 4–20mA.
Communication: RS232.
(Ethernet pending)
Sensor Drive Circuit: Proprietary
Sensor Drive Circuit provides
enhanced flow signal stability and insensitivity to process temperature
changes
Display: 2-line LCD back-lit Flow Rate
(top line),Total and Temperature (bottom line) is standard on all Mass
Flow Meters
(except for SIL and SRL Series which have no displays).
Menu Navigation: 4-Button Touch Display
Screen9 Keypad (except SIL & SRL) and RS232 with navigational software
(Sage VIP) standard (software optional on SIL & SRL).With Touch
Display Screen technology, the cover does not need to be opened to access
Menuing System.
Multiple Channel Capability: Up
to four totally independent calibrations available
on all products except SIL & SRL. Calibrate for
four different gases, different sensitivities, and/or
different configurations (Channels A–D). Channels
can be keypad, laptop or externally selectable (via
contact closures).
Power: 24VDC
(250 ma with back light off, 350mA max with
back light on), 115VAC or 230VAC (Switchable) on SIG & SRG Series.
250mA max on SIE, SRE, SIL or SRL Series
FLOW RANGE / SIZES
Units
of Measurement: Flow–SCFM, SCFH, SCFD, SCCM,
NCMM, NCMH, KG/S, KG/M, KG/H, KG/D, LBS/S, LBS/M, LBS/H,
LBS/D, SLPM, SLPH; Velocity– SFPM, SFPS,NMPM,NMPS,NMPH
(other combinations available); Temperature–°C and °F |
Insertion
Meters: Full Scale up to 35,000 SFPM (i.e.,
up to 12,000 SCFM in an 8" Sch. 40 Pipe). Higher velocities
optionally available. Resolve as low as 10 SFPM. Standard probe
is 1/2" diameter,with optional 3/4". Standard lengths are
6", 12", 15", 18" or 24" and optional up to
48"
In-Line Meters: Full Scale up to 3150
SCFM (4" Flow Meter).
Resolve as low as .006 SCFM (3/8" Flow Meter).Male NPT
fittings standard. 150# and 300# flanged ends optional.
(See chart below)
PERFORMANCE
Standard
Flow Accuracy6:+/-1% of
Reading +0.5% of Full Scale over 100: 1 turndown.(Contact
Sage for special accuracy)
Repeatability: 0.2%
of Full Scale
Turndown: 100:1
of Full Scale minimum (1000:1 with multiple channels)
Calibration: Sage
Metering’s National Institute of Standards
Traceable (NIST) calibration facility
Gas Temperature: Std.:
-40°F to 200°F (-40°C to 93°C );
HT01:
200°F to 350°F (93°C to 177°C ); HT02: 350°F to 500°F
(177°C to 260°C); HT03: 500°F to 750°F (260°C to
399°C)
Integral or Remote Enclosure
Temperature: 0° to 150°F (-18°C
to 65°C ). Contact Sage for lower temperature
ranges.
Note: Remote Enclosure can be mounted up to 2000 feet away to locate
in a suitable temperature environment
Pressure Rating: 500
psig (1000 psig optional)
Response Time: 1 second
(each time constant) for flow change
Subsystem Interface: Full
duplex digital using SLIP protocol
ENCLOSURES
Integral General Purpose, Integral Explosion Proof, Remote General Purpose, and Remote Explosion Proof. Remote styles do not require electronics at sensor Junction Box. (See dimensional drawings on page 5 for classifications.)
DISPLAY
/ KEYPAD
General Purpose SIG & SRG Transmitter have large format 2 x 16 LCD back-lit display with large format 4-Button Keypad. Explosion Proof SIE & SRE Transmitter has 2 x 16 small format back-lit display with standard 4-Button Keypad. (SIL & SRL transmitter are blind.) Keypad uses Touch Display Screen Technology on General Purpose Meters (SIG & SRG) and can be activated without removing cover.
|
WETTED
PARTS
316L Stainless Steel for Flow Bodies, Sensor Flow Elements and Flow Conditioners. Hastelloy (recommended for Chlorine Gas) and other materials optional.
FLOW CONDITIONING
All In-Line Flow Meters 3/4 inch and above include built-in flow conditioning
RELAYS
Two 1-amp relay channels (each SPDT) optionally available
on SIG & SRG
Series.One 1-amp relay is standard on SIE & SRE Series.Menu configurable
(see description under Sage Thermal Mass Flow Meters on page 3)
AGENCY APPROVALS
CSA and UL. Contact Sage or visit our website (www.sagemetering.com) for
details
ETHERNET
COMPATIBILITY
Pending: Ethernet 10Mb/s or ID base-T via RJ-45 Cat 5 cable. Contact Sage or
visit our website (www.sagemetering.com) for details.
LIMITED WARRANTY
Sage Metering's Series of Thermal Mass Flow Meters are warranted against faulty
materials or workmanship for one year from the date of delivery to the buyer.After
issuance of a Return Meter Authorization (RMA) by Sage, and upon receipt of the
defective meter, Sage will either repair or replace the defective meter at its
sole option and at no cost to the purchaser. |